8.8.2
Commands
────────────────────────────────────────────────────
■
Queries the line voltage monitored parameter.
Syntax
:MEASure:VOLTage?
Function
Returns the line voltage monitored parameter as a 4-digit numerical value
in NR3 format.
Note
Use this command only after using the ":MODE" command to set the
measuring mode.
Example
Transmission
:MEASure:VOLTage?
Response
Headers ON
:MEASURE:V99.90E+00
Headers OFF
+99.90E+00
The line voltage monitored parameter is 99.9 V.
Error
(ME equipment) (ordinary equipment)
If the measuring mode has not been set, or when in the low resistance
measurement mode, an error occurs.
■
Queries the measured value.
Syntax
:MEASure?
Function
Returns the measured value and its judgment as numerical values in NR3
and NR1 format.
The data are ordered as follows: <measured value> and <judgment>.
<Judgment>
(If the allowable value judgment is disabled, 0)
Allowable value or lower (PASS): 0
Higher than allowable value (FAIL): 1
Note
Use this command only after using the ":MODE" command to set the
measuring mode.
Example
Transmission
:MEASure?
Response
Headers ON
:M2.345E‑03,1
Headers OFF
+2.345E‑03,1
The measured value is 2.345 mA, and the allowable value judgment is
"higher than allowable value (FAIL)."
Error
(ME equipment) (ordinary equipment)
If the measuring mode has not been set, an error occurs.
:MEASure:VOLTage?
:MEASure?
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......