![Hioki 3155-01 Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164033.webp)
13
────────────────────────────────────────────────────
3.1 Measurement Networks
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
3.1.1 9497 NETWORK B (for the Medical Electrical Equipment)
V
1 k
Ω
±
0.1%
10 k
Ω±
0.1%
0.015
μ
F
±
1%
V
1 k
Ω±
0.1%
Network with frequency characteristics
(ON)
Network with only 1 k
Ω
of uninduced
resistance (OFF)
Chapter 3
Options
3.1 Measurement Networks
Please choose one of the three measurement networks that simulate the
human body to perform measurements conforming to standards for electrical
equipment to be tested.
The measurement networks cannot be used separately, but must always be
installed on the 3155.
For the installation procedure, refer to Section 4.1, "Installing the
Measurement Network," and for the fuse replacement, Section 10.2.2,
"Replacement of the Measurement Network Fuse."
This measurement network is for the medical electrical equipment.
Measurement network B conforms to the following standard:
Medical electrical equipment Part 1: General requirements of safety (IEC
60601-1 (1988-12) + am1 (1991-11) + am2 (1995-03))
By installing on the 3155 and making the appropriate settings from the
screen, a configuration with a network with frequency characteristics for ON,
and a network with only 1 k
Ω
of uninduced resistance for OFF is established.
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......