![Hioki 3155-01 Скачать руководство пользователя страница 272](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164272.webp)
252
────────────────────────────────────────────────────
8.8.2
Commands
────────────────────────────────────────────────────
■
Queries the condition of the equipment to be measured in the manual measurement.
Syntax
:CONFigure:CONDition?
Function
Returns the setting for the condition of the equipment to be measured as
<data>.
<data>
NORMAL
Normal condition
EARTH
Single fault condition (open ground)
POWERSOURCE Single fault condition (open power lead)
Note
Use this command only after using the ":MODE" command to set the
measuring mode.
Example
Transmission
:CONFigure:AUTO OFF;:CONFigure:CONDition?
Response
Headers ON
:CONFIGURE:CONDITION NORMAL
Headers OFF
NORMAL
The equipment to be measured has been set to the normal condition.
Error
(ME equipment) (ordinary equipment)
If the measuring mode has not been set, or when in the low resistance
measurement mode, or when in automatic measurement mode, an error
occurs.
■
Sets the contact condition in the enclosure leakage current mode.
Syntax
:CONFigure:CONTact_<
data>
data
EQUipment/EARTh
Function
Sets the contact condition in the enclosure leakage current mode.
EQUipment Sets to between parts of the enclosure.
EARTh Sets to between te enclosure and the earth.
Note
Use this command only after using the ":MODE" command to set the
measuring mode to the enclosure leakage current.
When the 9498 NETWORK C or the 9499 NETWORK D is installed, select
"EQUipment" to set to between the enclosure and the power line.
Example
Transmission
:CONFigure:CONTact EARTh
The contact condition is set to between the enclosure and the earth.
Error
(ME equipment) (ordinary equipment)
If the measuring mode has not been set, or when in a mode other than the
enclosure leakage current mode, an error occurs.
:CONFigure:CONDition?
:CONFigure:CONTact
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......