7.3 Enclosure Leakage Current Measurement Mode (Universal)
────────────────────────────────────────────────────
7.3.3 Connecting Equipment for Leakage Current Measurement
CAUTION
・
The maximum current that can be input into the leakage current
measurement terminals of the 3155 is 25 mA (AC+DC). The input of any
larger current will cause the beeper to sound (when the beeper is turned on),
after which continued input may cause the fuse in the measurement network
to fail. In this case, input must be terminated immediately.
Afterwards, a fuse check is performed automatically upon returning to the
initial screen from the leakage current measurement mode. Therefore, do
not connect anything to leakage current measurement terminal T1. (If
anything is connected, the display may show "Measurement Network Fuse
blown.")
・
The rated capacity of the auxiliary power socket is 1500 VA or less. If this
capacity is exceeded, the auxiliary power outlet circuit breaker will actuate,
and the power supply to the equipment to be measured will be cut off.
(With the 3155-01, the main power switch is turned off.)
・
If enclosure leakage current is to be measured between the enclosure and
the power supply line, or the power of the equipment to be measured
exceeds 1500VA, make the connection directly to the outlet; do not use the
auxiliary power socket. (When measuring class
equipment, connect the
protective earth conductor of the equipment to be measured to the G
terminal of the 3155.)
In this case, the power source polarity and the condition of the equipment to
be measured cannot be set on the 3155.
・
The table tap is subject to line voltage. Watch out for electrical shocks.
・
In cases where the voltage is applied directly to the leakage current
measurement terminals of the the 3155 when making a measurement, the
measured value may be subject to synchronous voltage. In which case, an
accurate measurement cannot be obtained without first performing zero
adjustment. Always perform zero adjustment before measuring.
(1) When setting to "between the enclosure and the power line."
No measurement is made for internally powered equipment.
1. On the enclosure leakage current measurement screen, set the contact
condition to "between the enclosure and the power line."
2. Plug the power cord of the equipment to be measured into a table tap.
3. Plug the table tap into an outlet.
4. Connect the red 9170 TEST LEAD to the leakage current measurement
terminal T2, and the black one to T1.
5. Perform zero adjustment. For details, refer to Section 7.3.4, "Zero
Adjustment (Universal - Enclosure Leakage Current)."
6. Plug the red 9170 TEST LEAD into the table tap.
7. Touch the black 9170 TEST LEAD to the enclosure of the equipment to be
measured.
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......