Commands
────────────────────────────────────────────────────
■
Sets the measurement network filter.
Syntax
:CONFigure:FILTer_<
data>
data
ON/ON1/ON2/OFF/OFF1/OFF2
Function
Sets the measurement network filter.
(ME equipment)
9497 NETWORK B
ON Sets to the network with frequency characteristics.
OFF Sets to the network with only 1 k
Ω
of uninduced resistance.
(Ordinary equipment)
9498 NETWORK C
ON1 Sets to the perception and reaction network.
ON2 Sets to the let-go network.
OFF Sets to the body impedance network.
9499 NETWORK D
OFF1 Sets to the network of 1 k
Ω
.
ON
Sets to the network of 1.5 k
Ω
and 0.15
μ
F.
OFF2 Sets to the network of 2 k
Ω
.
Note
Use this command only after using the ":MODE" command to set the
measuring mode.
Example
Transmission
:CONFigure:FILTer OFF
The measurement network filter is set to the network with only 1 k
Ω
of
uninduced resistance. (ME equipment)
Error
(ME equipment)
If the measuring mode has not been set, or when in the low resistance
measurement mode, an error occurs.
(Ordinary equipment)
If the measuring mode has not been set, or when in the low resistance
measurement mode, an error occurs.
When the 9498 NETWORK C is installed, and the measurement current is
AC peak, if set to OFF, an error occurs.
:CONFigure:FILTer
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......