5.4 Enclosure Leakage Current Measurement Mode (ME Equipment)
────────────────────────────────────────────────────
5.4.2 Connecting Equipment for Enclosure Leakage Current
Measurement (ME Equipment)
CAUTION
・
The maximum current that can be input into the leakage current
measurement terminals of the 3155 is 25 mA (AC+DC). The input of any
larger current will cause the beeper to sound (when the beeper is turned on),
after which continued input may cause the fuse in the measurement network
to fail. In this case, input must be terminated immediately.
Afterwards, a fuse check is performed automatically upon returning to the
initial screen from the leakage current measurement mode. Therefore, do
not connect anything to leakage current measurement terminal T1. (If
anything is connected, the display may show "Measurement Network Fuse
blown.")
・
The rated capacity of the auxiliary power socket is 1500 VA or less. If this
capacity is exceeded, the auxiliary power outlet circuit breaker will actuate,
and the power supply to the equipment to be measured will be cut off.
(With the 3155-01, the main power switch is turned off.)
・
When power of the equipment to be measured exceeds 1500VA, make the
connection directly to the outlet; do not use the auxiliary power socket.
(When measuring class
equipment, connect the protective earth conductor
of the equipment to be measured to the G terminal of the 3155.)
In this case, the power source polarity and the condition of the equipment to
be measured cannot be set on the 3155.
・
If the voltage application button to the 9196 APPLY UNIT is ON, the danger
of electrical shock is present. Turn the voltage application button OFF when
not making measurements.
・
Use an insulating transformer.
(1) Measurement between the enclosure and the earth. (With class
equipment,
between each part of the enclosure which is not protectively earthed and the
earth)
1. On the enclosure leakage current measurement screen, set the contact
condition to "between the enclosure and the earth."
2. Plug the power cord of the equipment to be measured into the auxiliary
power socket found on the right side of the 3155.
No connection is made for
internally powered equipment.
3. Connect the red 9170 TEST LEAD to the leakage current measurement
terminal T2.
4. Touch the red 9170 TEST LEAD to the enclosure of the equipment to be
measured.
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......