![Hioki 3155-01 Скачать руководство пользователя страница 321](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164321.webp)
301
────────────────────────────────────────────────────
9.3 Accuracy Tables
────────────────────────────────────────────────────
Range (accuracy range)
Resolution
Accuracy:
20 Hz
≦
f
≦
10 kHz
Accuracy:
DC
≦
f
<
20 Hz,
10 kHz
<
f
≦
1 MHz
25.00 mA (5 to 25 mA) *
5
10
μ
A
±
(1% rdg.+6dgt.
) ±
(2% rdg.+10 dgt.)
5.000 mA (500
μ
A to 5 mA) *
5,
*
6
1
μ
A
±
(1% rdg.+6dgt.
) ±
(2% rdg.+10 dgt.)
500.0
μ
A (40 to 500
μ
A) *
5,
*
6
0.1
μ
A
±
(1% rdg.+6dgt.
) ±
(2% rdg.+10 dgt.)
50.00
μ
A (4 to 50
μ
A) *
5,
*
6,
*
7
0.01
μ
A
±
(1% rdg.+6dgt.
) ±
(2% rdg.+10 dgt.)
Range (accuracy range)
Resolution
Accuracy
25.00 mA (5 to 25 mA)
10
μ
A
±
(0.2% rdg.+3 dgt.
)
5.000 mA (500
μ
A to 5 mA) *
6
1
μ
A
±
(0.2% rdg.+3 dgt.
)
500.0
μ
A (5 to 500
μ
A) *
6
0.1
μ
A
±
1.0% f.s.
50.00
μ
A (1 to 50
μ
A) *
6,
*
7
0.01
μ
A
±
1.0% f.s.
Range (accuracy range)
Resolution
Accuracy
75.0 mA (10 to 75 mA)
100
μ
A
±
(2% rdg.+2 dgt.
)
10.00 mA (1 to 10 mA) *
6
10
μ
A
±
(2% rdg.+2 dgt.
)
1.000 mA (100
μ
A to 1 mA) *
6
1
μ
A
±
2.5% f.s.
100.0
μ
A (10 to 100
μ
A) *
6,
*
7
0.1
μ
A
±
4% f.s.
Range
Resolution
Measured current
Accuracy
500.0
Ω
100 m
Ω
1 mA
±
(0.4% rdg.+5 dgt.
)
50.00
Ω
10 m
Ω
10 mA
±
(0.4% rdg.+5 dgt.
)
5.000
Ω
1 m
Ω
100 mA
±
(0.4% rdg.+5 dgt.
)
500.0 m
Ω
0.1 m
Ω
100 mA
±
0.5% f.s.
(3) Ranges for the 9499 NETWORK D
1. Measuring mode: AC/AC+DC
*
5
: In the AC measurement mode, frequency response of the high pass filter (fc
= 7 Hz) is added.
*
6
: Indicated ranges are for a resistance of 1 k
Ω
. For 1.5 k
Ω
and 2 k
Ω
, the
ranges are 1/1.5 and 1/2 times those indicated above.
*
7
: The setting cannot be made in the enclosure leakage current measurement
mode.
2. Measuring mode: DC
*
6
: Indicated ranges are for a resistance of 1 k
Ω
. For 1.5 k
Ω
and
2 k
Ω
, the ranges are 1/1.5 and 1/2 times those indicated above.
*
7
: The setting cannot be made in the enclosure leakage current
measurement mode.
3. Measuring mode: AC peak
*
6
: Indicated ranges are for a resistance of 1 k
Ω
. For 1.5 k
Ω
and
2 k
Ω
, the ranges are 1/1.5 and 1/2 times those indicated above.
*
7
: The setting cannot be made in the enclosure leakage current
measurement mode.
Accuracy of the low resistance meter
When measuring with 3157 measuring equipment settings, the accuracy of the
3157 AC GROUNDING HiTESTER is applied.
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......