![Hioki 3155-01 Скачать руководство пользователя страница 271](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164271.webp)
251
────────────────────────────────────────────────────
8.8.2
Commands
────────────────────────────────────────────────────
■
Sets the condition of the equipment to be measured in the manual measurement.
Syntax
:CONFigure:CONDition_<
data>
data
NORMal/EARTh/POWersource
Function
Sets the condition of the equipment to be measured in the manual
measurement.
NORMal Normal condition.
EARTh Single fault condition (open ground)
POWersource Single fault condition (open power lead)
Note
The setting cannot be made depending on the the condition of the
equipment to be measured setting and measuring mode.
For details, refer to the table below.
Use this command only after using the ":MODE" command to set the
measuring mode.
Example
Transmission
:CONFigure:AUTO OFF;
:CONFigure:CONDition NORMal
The condition of the equipment to be measured is set to the nomal
condition.
Error
(ME equipment) (ordinary equipment)
If the measuring mode has not been set, or when in the low resistance
measurement mode, or when in automatic measurement mode, or if the
setting cannot be made in the current condition, an error occurs.
Condition of the
equipment to be
measured
Measuring mode
Class
equipment
Class
equipment
Internally powered
equipment
N.C
S.F.C
N.C
S.F.C
N.C
S.F.C
Open
power
lead
Open
ground
Open
power
lead
Open
ground
Open
power
lead
Open
ground
Earth leakage current
Yes
Yes
No
−
−
−
−
−
−
Enclosure leakage current
Yes
Yes
Yes
Yes
Yes
No
Yes
No
No
Low resistance measurement
−
−
−
−
−
−
−
−
−
Patient auxiliary current
Yes
Yes
Yes
Yes
Yes
No
Yes
No
No
Patient leakage current
Yes
Yes
Yes
Yes
Yes
No
Yes
No
No
Patient leakage current
No
No
No
No
No
No
No
No
No
Patient leakage current
No
No
No
No
No
No
No
No
No
:CONFigure:CONDition
Conditions available by the "CONFigure:CONDition" command
N.C: Normal condition
S.F.C: Single fault condition
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......