5.6 Patient Leakage Current
Measurement Mode
────────────────────────────────────────────────────
9
Allowable value
Current setting of the allowable value (Single fault condition only)
10
Setting display for the
equipment to be measured
The following parameters are displayed:
Settings of the equipment to be measured
Model name, Number, Current consumption monitor value, Line voltage
monitor value
Key
Description of the function
11
(Clear the maximum
value)
Use as necessary to clear unnecessary data and start new measurements.
As long as this key is not pressed, the maximum value is not cleared even after
changing measurement settings in the same measuring mode.
The maximum value is cleared after changing the measuring mode.
12
Opens the allowable value set-up window. (Single fault condition only)
Range of the allowable value: 5
μ
A to 20 mA
For details, refer to Section 5.10.1, "Allowable Value Setting."
13
Opens the polarity and equipment condition set-up window.
1. Select either the
or
mode.
For automatic measurement, refer to Section 5.10.3, "Automatic
Measurement Setting."
2. When in the manual mode, in addition to making measurements while
changing the polarity and condition of the equipment to be measured, the
maximum value data can be printed out or saved.
Power source polarity selection:
NORMAL:
, REVERSE:
14
Opens the measurement network filter set-up window.
Select either filter ON or OFF.
: Network with frequency characteristics
: Network with uninducted resistance of only 1 k
Ω
15
Opens the print-out window.
Prints out the maximum value displayed on the screen.
For details, refer to Section 5.10.5, "Printing out the Maximum Values."
16
Opens the save data window.
For details, refer to Section 5.10.6, "Saving the Maximum Values."
17
Exits the patient leakage current
measurement, and returns to the initial
screen.
(2) Settings
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......