![Hioki 3155-01 Скачать руководство пользователя страница 183](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164183.webp)
163
────────────────────────────────────────────────────
6.6 Condition Set-up Window Setting (for IEC/TR 60990)
────────────────────────────────────────────────────
6.6.7 Low Resistance Measurement with 3157
Notes when saving data
・
Even if the filter setting, measurement current and contact condition have
changed, as long as the registered data unit is the same, the data for that
data unit will be overwritten. (after a request for confirmation)
・
When saving data manually for each setting condition (power source polarity
and condition of the equipment to be measured), clear the maximum value
prior to each measurement and save after changing the model name and
number. For automatic measurement, refer to Section 6.6.3, Automatic
Measurement Setting (for IEC 990)."
・
For changing the model name and number, refer to Section 6.2.2,
"Equipment Set-up Screen."
Settings
(1) On the measuring mode screen, press the
key.
(2) At the bottom of the screen, the save data window will
open.
Press the
key to save the maximum values.
If a data unit with the same name already exists, a
request for confirmation appears.
Press the
key to close the save data window
without saving.
If the
key is pressed on the measuring mode
screen without first either saving or printing out the
current data, that data will be lost.
Low resistance measurement mode allows measurement with optional 3157
AC GROUNDING HiTESTER (maximum 31 AAC, in accordance with multiple
standards). Using the RS-232C interface enable the 3155 to control the 3157.
Prior to set up, connect 3157 to equipment to be measured and 3157 to 3155
with connection cable. Synchronize 3157 communication settings with 3155.
3157 communication settings are applied when power is turned ON. Set up
and configure while power is OFF.
See "When using 3157" on the previous page.
See also Section 8.5.1, "Communication Conditions Setting."
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......