9.2 Measuring Mode Specifications
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
9.2.1 Leakage Current Measurement
Measurement method
A current value is calculated and displayed by measuring the voltage
drop across a simulated human body resistance (measurement network).
A/D conversion format
Sigma-delta format
Display update rate
5 times/second (moving average display)
Maximum value hold
Maximum leakage current value and power source condition
Permissible measured
current
25 mAAC/DC
Permissible common mode
voltage
(When using the measurement terminals T1 and T2)
250 VDC+AC peak (for ranges other than 50
μ
A)
20 VDC+AC peak (for the 50
μ
A range)
Measured currents
Earth leakage current
Enclosure leakage current
Patient leakage current
*
1
Patient leakage current
*
1,
*
2
Patient leakage current
*
1,
*
2
Patient auxiliary current*
1
(*
1
: measurable only when using the 9497 NETWORK B)
(*
2
: measurable when using the external power source such as the
9196 APPLY UNIT)
Power source polarity setting
Normal polarity / reverse polarity
Single fault condition setting
Open power lead / open ground
Measurement method setting
Manual / automatic (for a period from 1 to 99 seconds)
Measuring mode
AC / DC / AC+DC / AC peak
(Peak measurement can be made only when using the 9498 NETWORK
C or the 9499 NETWORK D.)
Measurement range
25 mA /5 mA / 500
μ
A / 50
μ
A
Range switching
Full automatic
Frequency characteristics
DC to 1 MHz (DC, AC+DC) / 7 Hz (-3 dB) to 1 MHz (AC)
Temparature coefficient
±
(0.04% rdg.+0.5 dgt.)/
℃
Input resistance
Measurement terminals T1 and T2: 1 M
Ω±
2% (with no measurement
network installed)
Between measurement terminals T1 and T2: 2 M
Ω±
2% (with no
measurement network installed)
Input capacity
terminals T1 and T2)
(with no measurement network installed)
CMRR
Between measurement terminals T1 and T2, and frame
(within the permissible common mode voltage)
50
μ
A range: 40 dB min. (at 1 MHz)
Ranges other than 50
μ
A: 40 dB min. (at 100 kHz)
9.2 Measuring Mode Specifications
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......