6.3 Enclosure Leakage Current Measurement Mode (for IEC/TR 60990)
────────────────────────────────────────────────────
NOTE
Wire
Table tap
Equipment to
be measured
Insulating transformer
Red 9170 TEST LEAD
Protective earth terminal
(2) Measurement of leakage current to a telecommunication network
1. On the enclosure leakage current measurement screen, set the contact
condition to "between the enclosure and the earth."
2. Plug the power cord of the equipment to be measured into a table tap.
3. Use wire to connect the outlet earth with the table tap.
4. Plug the table tap into an insulating transformer.
5. Connect the red 9170 TEST LEAD to the leakage current measurement
terminal T2.
6. Touch the red 9170 TEST LEAD to the connection for a telecommunication
network of the equipment to be measured.
・
To change the power source polarity, connect wire connecting the protective
earth terminal of the outlet and one pole of the table tap to the other pole.
・
To change the connection for a telecommunication network, change the red
9170 TEST LEAD connection.
・
In the enclosure leak current measuring mode, the 50
μ
A range is not used.
・
When testing class 0I or other older models, use (UL1015, AWG #18 or
equivalent) wire to connect the G terminal with the protective earth
terminal of the equipment to be measured.
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......