![+GF+ Signet 9950 Скачать руководство пользователя страница 82](http://html2.mh-extra.com/html/gf/signet-9950/signet-9950_operating-instructions-manual_3044293082.webp)
82
GF 9950-10 Six Channel Transmitter
Environmental
Ambient Temperature:
DC Power ...............-10 °C to 70 °C (14 °F to 158 °F)
AC Power ...............-10 °C to 60 °C (14 °F to 140 °F)
Storage Temp .........-15 °C to 70 °C (5 °F to 158 °F)
Relative Humidity........ 0 to 100% condensing for (front only);
0 to 95% non-condensing (rear panel)
Maximum Altitude .......4,000 m (13,123 ft)
Enclosure Rating .......... NEMA 4X/IP65 (front face only)
Installation Category Cat II
Pollution Degree 2
Performance Speci
fi
cations
System Accuracy
• Primarily dependent upon the sensor.
System Response
• Primarily dependent upon the sensor. Transmitter adds
a maximum of 150 ms processing delay to the sensor
electronics.
• Minimum update period is 100 ms.
• System response is tempered by the display rate, output
averaging and sensitivity feature.
Shipping Weights
Base Unit ....................0.63 kg (1.38 lb)
Relay Module..............0.19 kg (0.41 lb)
Power Requirements
DC (3-9950-10,
3-9950-11
) 24 VDC nominal (12 to 32 VDC,
±10% regulated). A UL60950-1 or
UL61010-1 certi
fi
ed power supply
must be used. Power supply shall
also be rated for operation at
4000 m altitude.
AC (3-9950-11).................. 100-240 VAC 50-60 Hz, 24 VA max
Maximum Current ...............200 mA (w/o optional relay module)*
500 mA
(with optional relay module)*
Current Loop ...................... 12 to 32 VDC, ±10% regulated
4 to 20 mA (30 mA max.)
Overvoltage protection ....... 48 V Transient Protection Device
(for
DC
ONLY)
Current limiting for circuit protection
Reverse-Voltage protection
*The current draw of the other modules and the sensors are minimal
Electrical Requirements
Power to Sensors
Voltage............................... +4.9 to 5.5 VDC @ 25 °C, regulated
Current............................... 30 mA Maximum
Short Circuit .......................Protected
Isolation ............................. Low voltage (< 48 V AC/DC)
Standards and Approvals
• CE, UL, CUL, WEEE , UKCA
• RoHS
Compliant
China RoHS (Go to www.gfps.com for details)
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules.
Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received,
including interference that may cause undesired operation.
• Manufactured under ISO 9001, ISO 14001 and ISO 45001
Speci
fi
cations
General
Input channels ............Six channels, programmable for
Digital (S
3
L) or frequency input
Outputs.......................Two 4 to 20 mA output in base unit,
Two or four 4 to 20 mA outputs in
optional modules,
Two or four programmable relay
outputs optional
Enclosure and Display
Case Material .............PBT
Window .......................Shatter-resistant glass
Keypad ....................... 4 buttons, injection-molded
silicone rubber seal
Display ........................Dot Matrix LCD
Update Rate ...............1 s
LCD Contrast ..............5 settings
Indicators .................... Two horizontal digital bar graphs,
four LED relay status indicators
Enclosure
Size
.....................¼
DIN
Mounting
Panel
...................
¼ DIN, ribbed on four sides for panel
mounting clip inside panel, silicon
gasket included
Wall
......................
Large enclosure (sold as an accessory)
Terminal Blocks
Pluggable screw type: use minimum 105 °C rated wire
Torque ratings
Power/Loop..........0.49 Nm (4.4 lb-in.)
Freq/S
3
L ...............0.49 Nm (4.4 lb-in.)
Relay....................0.49 Nm (4.4 lb-in.)
Connector wire gauge:
Power, Loop .........12 to 28 AWG
Freq/S
3
L ...............16 to 28 AWG
Module connector wire gauge:
Relay....................12 to 28 AWG
Sensor Input Speci
fi
cations
Digital (S
3
L) Sensors ....... Serial ASCII, TTL level, 9600 bps
Frequency (Flow) Sensors: 0.5 to 1200 Hz
Accuracy .......................... ± 0.5% of reading max error @ 25 °C
Resolution........................ 1
μ
s
Repeatability .................... ± 0.2% of reading
Rejection.......................... No Effect ± 1
μ
A per volt
Short Circuit ..................... Protected
Reverse Polarity .............. Protected
Update Rate .................... (1/frequency) + 100 ms
Graphical User Interface created with emWin licensed by SEGGER