40
GF 9950-10 Six Channel Transmitter
CAL Menu
ON prevents relays from activating while making adjustments, and relays in PULSE mode will
suspend pulsing. Output is held until the user exits the CAL menu or turns it OFF.
Press the
►
arrow, using the
▲
or
▼
to select OFF/ON. Press ENTER to save.
Default = OFF
Manually set value to match to a known liquid standard (external reference, NIST solutions).
Manually set to apply a linear offset to the Temperature measurement element.
SAVING will appear if offset is acceptable, VALUE TOO HIGH will appear if offset is outside of
range.
Manually enter the known Salinity value of your speci
fi
c Ocean.
NOTE: Performing a Salinity calibration does not affect the Conductivity calibration.
WARNING: Salinity calibration is not necessary if a standard conductivity calibration is performed and
the customer is satis
fi
ed with the results.
Press
►
to reset the Conductivity calibration to factory default.
Press
►
to reset Temperature calibration to factory default.
Press
►
to reset the Salinity calibration to factory default.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
CH 2 SALINITY
HOLD OUTPUTS
OFF
CH 2 SALINITY
C
SALINITY
S
SET
CH 2 SALINITY
TEMPERATURE
SET
CH 2 SALINITY
RESET COND CAL
CH 2 SALINITY
RESET TEMP CAL
°NF
CH 2 SALINITY
12-31-2015 XX
AVER
LAST CAL
AGE
CH 2 SALINITY
C
CONDUCTIVITY
S
SET
CH 2 SALINITY
RESET SAL CAL
°NF
Salinity