GF 9950-10 Six Channel Transmitter
Press ENTER to start assigning sensors to each channel.
Use the
▲
and
▼
arrow to go to any of the six channels directly.
NOTE: Display only available when the 9950 is in FACTORY mode.
Press the
►
arrow, and use the
▲
and
▼
arrow to select which TYPE of the sensor will be assigned
to the channel(s) (1-6).
Choose pH, and press ENTER to save sensor type.
Press the
►
arrow, using the
▲
or
▼
to select the input source to the channel.
Sources list will only contain pH sensors electronics (2751-X)
To change sensor selection after its been assigned, change the source to NONE, then the sensors
can be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered.
Enter 17-character string.
Default = pH
Select °F or °C.
Default = Determined by the SELECT UNITS screen menu. Metric = °C, U.S. Customary = °F.
Dampens display, output and relay response rates.
Select Low, Med, High, OFF. Default = OFF
NOTE:
GF Strongly recommends leaving Averaging OFF for pH, ORP, and Pressure measurements.
2751, pH Only. Enter the time between automatic impedance measurement updates in 15 minute
increments, 0 = Off - manual updates only.
Default = 60
2751, pH Only. Enter the impedance value for Broken pH Glass, 0 = Off, Default = 3
2751, pH only. Enter the high Impedance warning set point as NONE, 2, 3, 4, or 5 times the factory
Impedance of the electrode.
Default = NONE