35
GF 9950-10 Six Channel Transmitter
INPUT Menu
Press ENTER to start assigning sensors to each channel.
Use the
▲
and
▼
arrow to go to any of the six channels directly.
NOTE: Display only available when the 9950 is in FACTORY mode.
Press the
►
arrow, use the
▲
and
▼
arrow to select which TYPE of sensor will be assigned to the
channel(s) (1-6).
Choose TEMP, press ENTER to save sensor type.
Press the
►
arrow, using the
▲
or
▼
to select the input source to the channel.
Sources list will only contain TEMPERATURE sensors or modules.
To change sensor selection after it's been assigned, change the source to NONE, then the sensors can
be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered.
Enter 17-character string.
Default = TEMPERATURE
Select °F or °C.
Default = Determined by the SELECT UNITS screen menu. Metric = °C, U.S. Customary = °F.
Dampens display, output and relay response rates.
Select Low, Med, High, OFF.
Default = OFF
CHANNEL 1
TEMPERATURE
NAME
F
CH1 TEMPERATURE
TEMP UNITS
AV
°F
FE°°
CH1 TEMPERATURE
OFF
AVERAGE
SET UP
CHANNEL 1
CH 1 FACTORY
FACTORY
TYPE
CH1 TEMPERATURE
NONE
SOURCE
ENTER
Temperature
CAL Menu
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu or turns it OFF.
Press the
►
arrow, using the
▲
or
▼
to select OFF/ON. Press ENTER to save.
Default = OFF
Provides a maximum 20 °C offset to match to a known standard (external reference).
Resets Temperature Calibration to factory settings. After pressing
►
, select YES/NO.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
CH1 TEMPERATURE
HOLD OUTPUTS
OFF
CH1 TEMPERATURE
C
SET TEMP->
NT
CH1 TEMPERATURE
RESET CAL->
T
CH1 TEMPERATURE
12-31-2015 XX
LAST CAL