GF 9950-10 Six Channel Transmitter
Binary Input
RELAY Menu
(ALL) Select source for each Relay 1 - 4 (relay base unit) or RELAY A, B, C and D (external 8059 relay
module). Choose PRIMARY (CH 1-6), SECONDARY (CH 1-6), BINARY 1 - 4, DERIVED FUNCTION 1 - 4,
TIMER, or MULTIPLE. Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity
and Dissolved Oxygen. Derived Functions are only available when Derived Functions (CH 1-4) are enabled in
the OPTIONS menu.
(ALL) Select the desired mode of operation to switch the Relay state (ON / OFF). Modes avaliable OFF, LOW,
HIGH, WINDow IN, WINDow OUT, PROP PULSE, VOL PULSE, PWM, TOTAL, USP and ERROR. (Refer to
page 18, Relay Modes). Default = OFF. When MODE is set to ERROR, relay energizes if sensor problem is
detected on any channel. ON DELAY delays the energizing of the relay until after the programmed delay time.
(All Binary) Binary Inputs can be set to active the relay.
Requires the 3-9950.393-3 module for use.
(All Binary) Binary Inputs can be set to enable the relay based on whether the Binary input is ON (Closed) or
OFF (Open).
Requires the 3-9950.393-3 module for use.
English, French, German, and Spanish only.
A custom label can be assigned to the relay for ease of identi
fi
cation.
(ALL) (Shown if LOW, HIGH, WIND IN/OUT, or ERROR mode) Set in seconds (up to 9999.9) to delay
activating the relay once the relay condition is true.
Select YES to have the Red Backlight illuminate during relay activation.
Select NO to not have the Red Backlight activated by the relay activation.
Default backlight setting - NO
(ALL) Press
▲
or
▼
to turn relay ON or OFF for testing purposes.
Can also be used to reset or latch/unlatch the relay.
Does NOT reset the Totalizer.
RELAY 1
BINARY 1
SOURCE
RELAY 1
OFF
MODE
RELAY 1
RELAY 1
LABEL
R1 ON DELAY
SEC
0.00
RELAY 1 ACTIVATE
m
NO
A
RED BACKLIGHT
RELAY 1
mA
TEST RELAY
When the relay source is set to BINARY. The Binary Inputs
can sense if the input is ON (Closed) or OFF (Open).
The four binary inputs only control relay activation and are not
considered a measurement source.
Each binary mode is independent and each of them can be
assigned to a relay mode.
•
Binary Input from any point level switches to the 9950 allows
controlling a relay activity.
•
The 9950.393-3 relay module allows up to four binary (point
level sensors or similar) to be used to assist in relay control
of an application.
•
Binary Input signals can be used with the MULTIPLE source
input menu in the relay menu.
Relay source display allows operator to scroll thru relay 1 -4 on base unit and A-D on the external relay module.
External relays that are located in the 8059 External Relay module are limited to the basic relay modes as shown in the chart.
RELAY 1
CH 1 PRIMARY
SOURCE
RELAY 1
OFF
MODE