30
GF 9950-10 Six Channel Transmitter
View Measurement 2 displays the primary value and the channel name. This screen does not time out.
PRESSURE Setup Checklist
1. Make sure PRESSURE sensor type is selected.
(Refer to System Setup Menu, page 11)
2. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
3. Set Units of Measurement (PSI, BAR, KPa).
4. Set Last Cal Date and initials.
5. If applicable, set up relay functions for your application.
Displays the pressure reading. This
is the normal display and does not
time out.
Pressure
VIEW Mode Menu
CH1
7.0
77.0
4.0
99.0
CH2
PSI
°F
PSI
°F
10.0
0.0
10.0
0.0
CH1
90.0
PRESSURE
CH2
89.0
PRESSURE
PSI
PSI
10.0
0.0
10.0
0.0
Press ENTER to start assigning sensors to each channel.
Use the
▲
and
▼
arrow to go to any of the six channels directly.
NOTE: Display only available when the 9950 is in FACTORY mode.
Press the
►
arrow, use the
▲
and
▼
arrow to select which TYPE of sensor will be assigned to the
channel(s) (1-6).
Choose PRESSURE, press ENTER to save sensor type.
Press the
►
arrow, using the
▲
or
▼
to select the input source to the channel.
Sources list will only contain PRESSURE sensors or modules.
To change sensor selection after it's been assigned, change the source to NONE, then the sensors can be
reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered.
Enter 17-character string.
Default = PRESSURE
Press the
►
arrow, using the
▲
or
▼
to select the units of pressure measurement - PSI, BAR, or KPa.
Default = Determined by the SELECT UNITS screen menu.
Metric = Bar, U.S. Customary = PSI.
Dampens display, output and relay response rates.
Select: Low, Med, High, OFF.
Default = OFF. GF strongly recommends leaving averaging OFF for pH and pressure measurements
CHANNEL 2
PRESSURE
NAME
CH 2 PRESSURE
PSI
CH 2 PRESSURE
CUST CE
OFF
L CONST
AVERAGE
PRESS UNITS
SET UP
CHANNEL 1
CH 1 FACTORY
TYPE
CH 1 PRESSURE
CUST CE
NONE
L CONST
SOURCE
FACTORY
ENTER
INPUT Menu