GF 9950-10 Six Channel Transmitter
Level / Volume
CAL Menu
ON prevents relays from activating while making adjustments, and relays in PULSE mode will
suspend pulsing. Output is held until the user exits the CAL menu or turns it OFF.
Press the
►
arrow, using the
▲
or
▼
to select OFF/ON. Press ENTER to save.
Default = OFF
Shows SET LEVEL on middle line. When user presses any key, the live value is frozen and the user
edits that value. Returns either GOOD CAL or LEVEL OFFSET TOO LARGE.
Resets calibration to factory default. After pressing
►
, select YES/NO.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
CH 2 LEVEL
HOLD OUTPUTS
OFF
CH 2 LEVEL
CH 2 LEVEL
REU
RESET CAL
ELL
CH 2 LEVEL
pH STANDARD ->
SET
LEVEL CAL
CH2 LEVEL
12-31-2015 XX
LAST CAL
TEMPERATURE Setup Checklist
1. Make sure TEMPERATURE sensor type is selected.
(Refer to System Setup Menu, page 11)
2. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
3. Set Units of Measurement (°C or °F).
4. Set Last Cal Date and initials.
5. If applicable, set up relay functions for your application.
Displays the temperature value.
This is the normal display and does
not time out.
View Measurement 3 displays the primary value and the channel name. This screen does not time out.
Temperature
CH1
113.0
9999.9
89.0
9999.9
CH2
°F
GPMS
°F
0.00
100.00
0.00
100.00