GF 9950-10 Six Channel Transmitter
pH
pH Setup Checklist
1. Make sure pH sensor type is selected.
(Refer to System Setup Menu, page 11)
2. Set the Temperature Units (°C or °F).
3. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
4. Perform calibration (EasyCal, Standard or Standard and Slope).
5. Set Last Cal Date and initials.
6. Select source for Relay output (pH or Temp).
7. If applicable, set up relay functions for your application.
Displays the pH value and the
temperature below. This is the normal
display and does not time out.
15.00
15.00
CH1
7.0
77.0
4.0
99.0
CH2
pH
°F
pH
°F
-1.00
-1.00
VIEW Mode Menu
View Measurement 2 displays the primary value and the channel name. This screen does not time out.
Displays the RAW millivolt input from the electrode. Use this display to determine the relative
condition of your electrode during periodic calibration.
[Glass Impedance] 2751 only, pH only, last measured glass Impedance, press
►
key to manually
update reading.
2751-X, 274X electrodes Only, last reference impedance measurement, press
►
key to update
reading.
2571-X only. Using the
►
key, you will see the following screens: [Serial Number] Electrode Serial
Number, [Model Number] Electrode Model Number, [pH Calibration Slope] Calibration Slope Value,
[pH Offset] Calibration Offset, [Temperature Offset] pH only, Temperature Offset Value, [Factory
Impedance] pH only, Factory pH glass Impedance, [Usage Time] Electrode Runtime in Hours,
[Minimum pH] Minimum Measured Value, [Maximum pH] Maximum Measured Value, [Minimum
Temperature] pH only, Minimum Measured Temperature, [Maximum Temperature] pH only, <Maximum
Measured Temperature.
pH EASYCAL looks for buffers of 4, 7, or 10 pH buffers. (Refer to page 71)
CH1
7.0
pH
CH2
7.0
pH
15.00
15.00
pH
pH
-1.00
-1.00
RAW
CH2 182.10 mV
CH 1 264.10 mV
GLASS IMPEDANCE
CH1 1049 MOhm ->
EASYCAL
CH2 pH ->
CH1 pH ->
Sensor Data
CH1->
REF IMPEDANCE
CH1 27 MOhm ->