GF 9950-10 Six Channel Transmitter
The 9950 requires regulated
12 to 32 VDC,
±10% regulated
(24 VDC nominal) or 100-
240 VAC at 50-60 Hz
from an
external power supply
(not supplied).
Maximum current draw is:
500 mA with DC power
24 VA with AC power
Terminal Identi
fi
cation
Start-Up Guide
Prepare the transmitter installation location. If the back of the transmitter is dif
fi
cult to access when
installed, wire the removable terminal blocks
fi
rst, then install it completely.
+V
FREQ
DATA
GND
+V
FREQ
DATA
GND
Loop 1+
Loop 1–
Loop 2+
Loop 2–
PWR+
PWR–
CH2
CH1
USB
RELAY MODULE
MODULE 1
MODULE 2
Made in U.S. from U.S. and imported parts
N
L
100-240 V
A
C~
11-24 VDC
DC Power Terminals
Current Loop Outputs
1 and 2
Sensor Input 1
(Frequency/S
3
L)
AC Power Terminals
Sensor Input 2
(Frequency/S
3
L)
DC Power Terminal 3-9950-10 or 3-9950-11
Required by the instrument
• 12 to 32 VDC, ±10% regulated, 0.5 A MAX. DC power input and Loop current regulated voltage require the use of a
UL60950-1 or UL61010-1 certi
fi
ed power supply. Power supply shall also be rated for operation at 4000 m altitude.
AC Power Terminal 3-9950-11 Only
Required by the instrument
• 100-240 VAC at 50-60 Hz, 24 VA MAX
Relay Module
Dependent on model
• 3-9950.393-1 Four mechanical relays rated at 5A, 250 VAC or 5A, 30 VDC
• 3-9950.393-2 Two mechanical relays, 5A 250 VAC or 30 VDC, and two solid state relays rated at 50 mA, 30 VAC or 30 VDC
• 3-9950.393-3 Two mechanical relays, 5A 250 VAC or 30 VDC, and four binary inputs rated at 6 mA, 10 VDC to 24 VDC
Modules 1 and 2
• 3-9950.394-1 Single Channel Direct Conductivity/Resistivity Module
• 3-9950.394-2 Dual channel Conductivity Module
• 3-9950.395-M Modbus Module
• 3-9950.398-2 Dual Channel 4 to 20 mA Current Loop Output Module
Sensor 1 and Sensor 2: Digital (S
3
L)/Frequency Inpu
t
• V+:
+5 VDC out to sensor (black wire)
• FREQ:
Frequency input signal from sensor (red wire)
• S
3
L DATA:
Digital input signal from sensor (red wire)
• GND:
Sensor ground (white wire or silver wire from paddle wheel
fl
ow sensor)
Loop Outputs
• Two Passive 4 to 20 mA current loop outputs 12 - 32 VDC, ±10% regulated (30 mA Max)
• Up to 4 additional Loop Outputs via the Dual Channel 4 to 20 mA Output Module
ModBus Module
• Six Channel Primary and Secondary values