28
GF 9950-10 Six Channel Transmitter
Press ENTER to start assigning sensors to each channel.
Use the
▲
and
▼
arrow to go to any of the six channels directly.
NOTE: Display only available when the 9950 is in FACTORY mode.
Press the
►
arrow, use the
▲
and
▼
arrow to select which TYPE of sensor will be assigned to the
channel(s) (1-6).
Choose COND/RES, press ENTER to save sensor type.
Press the
►
arrow, using the
▲
or
▼
to select the input source to the channel.
Sources list will only contain Conductivity/Resistivity sensors or modules.
To change sensor selection after it's been assigned, change the source to NONE, then the sensors can
be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered.
Enter 17-character string.
Default = COND/RES
Press the
►
arrow, using the
▲
or
▼
to select the cell constant - 20.0, 10.0, 1.0, 0.1, 0.01, or CUSTOM.
Default = 1.0 (See NOTE below)
Display visible only when CUST is selected in the CELL CONSTANT selection of the menu.
Press the
►
arrow, using the
▲
or
▼
to enter the precise cell constant from the certi
fi
cate provided with
your sensor or from the information label on the sensor if CUST (custom) cell constant is selected.
Press the
►
arrow, using the
▲
or
▼
to select the measurement units
μ
S, mS, PPM, PPB, KOhm, or
MOhm.
Default =
μ
S
NOTE: In USP Relay Mode, TEMP COMP must be set to NONE and Unit Of Measure must be set to
μ
S.
View Measurement 2 displays the primary value and the channel name. The screen does not time out.
CH1
1205.00
77.0
987.9
65.5
CH2
uS
°F
uS
°F
100.00
0.0000
0.0000
100.00
Cond/Res Setup Checklist
1. Make sure COND/RES sensor type is selected.
(Refer to System Setup Menu, page 11)
2. Set Cell Constant.
3. Set the Temperature Units (°C or °F).
4. Set Conductivity units.
5. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
6. Set Temperature Compensation.
7. Set Last Cal Date and initials.
8. Select source for Relay output (COND or TEMP).
9. If applicable, setup relay functions for your application.
Displays the conductivity value and the
temperature below. This is the normal
display and does not time out.
Conductivity / Resistivity
CH1
1205.00
COND/RES
CH2
9999.9
COND/RES
uS
uS
0.0000
100.00
0.0000
100.00
VIEW Mode Menu
INPUT Menu
CH1 FACTORY
FACTORY
TYPE
CH1 COND/RES
NONE
SOURCE
SETUP
CHANNEL 1
ENTER
CHANNEL 2
COND/RES
NAME
CH 2 COND/RES
CELL CONSTANT
SENSOR 1.0
CH 2 COND/RES
CUST CELL CONST
SENSOR 1.00000
CH 2 COND/RES
COND UNITS
uS