42
GF 9950-10 Six Channel Transmitter
CAL Menu
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu or turns it OFF.
Press the
►
arrow, using the
▲
or
▼
to select OFF/ON. Press ENTER to save.
Default = OFF
Allows user to initiate the calibration process.
NOTE
: Dissolved Oxygen sensors are calibrated at the factory and do not require regular calibration.
Press
►
to begin the calibration process. User will be prompted to place sensor in 100% Solution
standard. Press ENTER to save value and establish a calibration point. If a single point calibration is
desired press the
▲
and
▼
keys to exit.
NOTE
: 100% calibration must be performed
fi
rst to allow access to the 0% calibration screen.
Dissolved Oxygen sensors are calibrated at the factory and do not require regular calibration.
Press
►
to begin the calibration process. User will be prompted to place sensor in 0% Solution standard.
Press ENTER to save value and establish a calibration point.
Resets Dissolved Oxygen calibration to factory settings. After pressing
►
, select YES/NO.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
CH 2 DO
HOLD OUTPUTS
OFF
CH 2 DO
C
100 % SOLUTION
S
SET ->
CH 2 DO
C
0 % SOLUTION
L
SET ->
CH 2 DO
RESET DO CAL ->
CHANNEL 2
12-31-2015 XX
°N
LAST CAL
F
Manually set Barometric value to match application altitude above or below sea level (506.62 - 1114.7 mBAR).
Default = 1013.2 (sea level)
Select °F or °C.
Default = Determined by the SELECT UNITS screen menu. Metric = °C, U.S. Customary = °F.
Dampens display, output and relay response rates.
Select Low, Med, High or OFF.
Default = OFF
INPUT Menu
Dissolved Oxygen
CH 2 DO
BAROMETRIC
1013.20
CH 2 DO
TEMP UNITS
°F
CH 2 DO
C
OFF
L CONST
AVERAGE