GF 9950-10 Six Channel Transmitter
ON prevents relays from activating while making adjustments and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu or turns it off.
Press the
►
arrow, using the
▲
or
▼
to select OFF/ON. Press ENTER to save.
Default = OFF
Set K-Factor (pulses per unit volume) from Flow Sensor manual.
Min: 0.0001, max. 9999999. Cannot be zero. Default = 60.0000
Sets the volume of each count of the Totalizer as a multiple of the volume unit of the K-Factor.
Min: 0.0001, max. 9999999. Cannot be zero. Default = 1.0000
Select to calibrate using rate method. (Refer to page 76)
Select to calibrate using Volume method. (Refer to page 76)
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
CAL Menu
CH 1 FLOW
HOLD OUTPUTS
OFF
CH 1 FLOW
MULTIPLE
RATE CAL ->
CH 1 FLOW
OFF
VOLUME CAL ->
CH1 FLOW
12-31-2015 XX
LAST CAL ->
CH 1 FLOW
60.00
KF
CH 1 FLOW
1.00
TF
(LABEL only) Identi
fi
es the totalizer units, has no effect on any calculations.
Default = LITERS
Set the Totalizer displayed in view mode. Select R TOTALIZER for resettable or select PERM TOTALIZER
for permanent.
Default = R TOTALIZER
The Sensitivity setting determines how the 9950 responds to sudden changes in the
fl
ow rate for
a frequency sensor. The value is expressed in units of measurement. If the setting is exceeded, it
"overrides" the Averaging function brie
fl
y to allow for the actual change in
fl
ow rate to be displayed.
Averaging resumes shortly after. The result is a smooth
fl
ow display and a quick response to large shifts in
the
fl
ow rate. To disable the sensitivity function set it to 99999.
Dampens display, output and relay response rates.
Select Low, Med, High, OFF.
Default = OFF
CH1 FLOW
OFF
AVERAGE
TOTALIZER
GALLONS
TOTAL UNITS
Flow
CH1 FLOW
SENSITIVITY
100.0
TOTALIZER
R TOTALIZER
DISPLAY
INPUT Menu