GF 9950-10 Six Channel Transmitter
View Measurement 2 displays the primary value along with the channel name.
The screen does not time out.
Displays the RAW millivolt input from the electrode.
Use this display to determine the relative condition of your electrode during periodic calibration.
2751, 276x electrodes only, last reference impedance measurement, press
►
to update reading.
2571 only. Using the
►
key, you will see the following screens: [Serial Number] Electrode Serial
Number, [Model Number] Electrode Model Number, [ORP Calibration Slope] Calibration Slope Value,
[ORP Calibration Offset] Calibration Offset, [Usage Time] Electrode Runtime in Hours, [Minimum
ORP] Minimum Measured Value, [Maximum ORP] Maximum Measured Value.
ORP EASYCAL looks for buffer of 87, 264, or 476 mV.
Press
►
to begin EASYCAL.
(Refer to page 73)
CH1
84.0
CUSTOM LABEL
CH2
74.0
CUSTOM LABEL
mV
mV
1000.0
1000.0
-999.0
-999.0
ORP Setup Checklist
1. Make sure ORP sensor type is selected
(Refer to System Setup Menu, page 11).
2. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
3. Set Averaging.
4. Perform calibration or set Standard (and Slope if desired).
5. Set Last Cal Date and initials.
6. If applicable, set up relay functions for your application.
Displays the ORP millivolt value.
This is the normal display and does
not time out.
ORP
CH1
76.0
ORP
CH2
50.0
ORP
mV
mV
1000.0
1000.0
-999.0
-999.0
RAW
CH2 182.10 mV
CH 1 264.10 mV
EASYCAL
CH2 ORP ->
CH1 ORP ->
VIEW Mode Menu
Sensor Data
CH1->
REF IMPEDANCE
CH1 27 MOhm ->