GF 9950-10 Six Channel Transmitter
NOTE:
Menu and Mode displays shown are examples only.
Your displays may vary.
Common Menus
The menu system shares certain modes between sensor
types.
The following describes the menus found in common between
most sensor types.
Common Menus
LOOP Menu
Select source for each loop. Choose PRIMARY, SECONDARY (CH 1-6) or DERIVED FUNCTION 1 - 4.
Secondary Values are only available for pH, Conductivity/Resistivity, Level, Salinity, and Dissolved Oxygen.
Derived Functions are only available for con
fi
gured Derived Functions.
English, French, German, and Spanish languages only.
A custom label can be assigned to the loop for ease of identi
fi
cation.
(COND/RES only) Select LIN/LOG. Default = LIN
Display only visible when Conductivity or Resistivity sensors are available.
Refer to LOG Current LOOP Output, page 66.
(ALL) Set value corresponding to desired 4 mA output. 6 digits max.
Default = 0 (ORP = -999).
(ALL) (Not shown in COND/RES LOG Mode)
Set value corresponding to desired 20 mA output. 6 digits max.
Defaults = 100 (Flow, Cond/Res, Temp), 14 (pH), 1000 (ORP), 10 (Lvl/Prs), 5 (4 to 20 mA), 42 (Sal).
(ALL) Set desired LOOP output value when sensor error (e.g., bad sensor, broken wire) is detected.
Select (3.6 mA, 22 mA,
or NONE
).
Default = 22
(ALL) Allows
fi
ne-tuning to compensate for errors in other equipment connected to the 9950.
Adjust the current output at 4 mA. The display value represents the precise current output.
Adjustment limits: from 3.80 mA minimum to 5.00 mA maximum. Default = 4.00 mA
(ALL) Allows
fi
ne-tuning to compensate for errors in other equipment connected to the 9900.
Adjust the current output at 20 mA. The display value represents the precise current output.
Adjustment limits: from 19.00 mA minimum to 21.00 mA maximum. Default = 20 mA.
(ALL) Press
▲
or
▼
to manually control the output current value from 3.8 mA to 21.00 mA to test the output of
LOOP. The current loop will revert to automatic control when this menu item is left.
(ALL) Manually select Sensor Type. Source to be assigned to the speci
fi
c loop output.
(Refer to page 11 for further instruction).
The following can individually be set for each loop on the 9950. Loops 1 and 2 are in the 9950 base unit.
Loops MOD1 A and MOD1 B are present when a loop module is in the Module 1 Slot. Loops MOD2 A and
MOD2 B are present when a loop module is in the Module 2 Slot.
LOOP 1
LOOP 1
LABEL
LOOP 1
LOG
MODE
L1 4 mA SETPOINT
0.00
UNITS
L1 20 mA SETPOINT
100.00
UNITS
LOOP 1
22 mA
ERROR VALUE
L1 ADJUST 4 mA
mA
4.00
L1 ADJUST 20 mA
mA
21.0
L1 TEST LOOP
mA
12.05
LOOP 1
CH SECONDARY
SOURCE