GF 9950-10 Six Channel Transmitter
Setting to ON prevents relays from activating while making adjustments or calibrating.
Relays in PULSE mode will suspend pulsing. 4 to 20 mA outputs are held until the user exits the CAL
menu or turns it OFF. Press the
►
arrow, using the
▲
or
▼
to select OFF/ON. Press ENTER to save.
Default = OFF
Select AT SENSOR to perform calibration using the GF 2751-2 Sensor Electronics.
Select AT INSTRUMENT to perform calibrations at the 9950 using the 3-2751-1, 3-2751-3, or 3-2751-4
electronics. (Refer to pH Calibration procedures, page 71).
Default = AT INSTRUMENT
(CAL AT INSTRUMENT only) Press
►
to start the EasyCal process.
The password prompt is the same as the other parameters.
(Refer to pH EasyCal procedure, page 71)
(CAL AT INSTRUMENT only) Applies a linear offset to the temperature measurement.
"SAVING" will appear if offset is acceptable, "ERR TOO LARGE TO CALIBRATE" if offset is outside of
range.
(CAL AT INSTRUMENT only) Applies an offset to the pH measurement. The slope value and the
standard value must be at least 2 pH units apart. The ideal values are the minimum and maximum
values of your process. (Refer to pH Calibration procedures, page 71).
Shows error message if offset is too high.
(CAL AT INSTRUMENT only) Applies a slope to the pH measurement. The slope value and the
standard value must be at least 2 pH units apart. The ideal values are the minimum and maximum
values of your process. (Refer to pH Calibration procedures, page 71).
Shows error message if slope is too low or high.
(CAL AT INSTRUMENT only) Press
►
to reset pH calibration to factory default.
(CAL AT INSTRUMENT only) Press
►
to reset temperature calibration to factory default.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
CAL Menu
CH2 pH
pH SLOPE
SET
CH2 pH
RESET pH CAL ->
SET
CH2 pH
RESET TEMPCAL ->
100.0
CH 2 pH
12 - 31 - 2015 XX
100
LAST CAL
.00
pH
CH 2 pH
EASYCAL ->
CH 2 pH
TEMPERATURE ->
SET
CH 2 pH
pH STANDARD ->
°
SET
F
CH 2 pH
HOLD OUTPUTS
OFF
CH 2 pH
AT INSTRUMENT
CAL