29
GF 9950-10 Six Channel Transmitter
Display only visible when PPM or PPB units are selected. Press the
►
arrow, using the
▲
or
▼
to select
the TDS factor, a TDS factor for converting the conductivity units to TDS is required.
Default = 0.50
Select °F or °C.
Default = Determined by the SELECT UNITS screen menu. Metric = °C, U.S. Customary = °F.
Dampens display, output and relay response rates.
Select Low, Med, High, OFF.
Default = OFF
Press the
►
arrow, using the
▲
or
▼
to select the temperature compensation (NONE, LINEAR, PURE
H2O). Default = LINEAR
NOTE: In USP Relay Mode in Conductivity, Relay Source must be set to COND, TEMP COMP must be
set to NONE and Unit Of Measure must be set to
μ
S.
Adjusted the % of temperature compensation.
Maximum slope setting is 9.99 % per ºC. Default = 2.0
(If Temperature Compensation setting is NONE, this item will not be displayed)
CH 2 COND/RES
TDS FACTOR
0.50
CH2 COND/RES
TEMP UNITS
°F
CH2 COND/RES
OFF
AVERAGE
CH2 COND/RES
LINEAR
TEMP COMP
CH 2 COND/RES
ADJ TEMP COMP
2.0
Conductivity / Resistivity
INPUT Menu
CAL Menu
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu or turns it OFF.
Press the
►
arrow, using the
▲
or
▼
to select OFF/ON. Press ENTER to save.
Default = OFF
Shows real-time value and automatically selects closest standard. “PLACE SENSOR IN STANDARD”.
Unit waits until reading is stable; if bad calibration, returns “ERROR, CANNOT DETERMINE
STANDARD”. Refer to buffer values and AUTO CAL Procedure. Refer to page 75.
Shows the live CONDUCTIVITY value on bottom line; when user presses any button the live value is
frozen and the user edits that value. If bad calibration, returns “ERR TOO LARGE TO CALIBRATE”.
Refer to Manual Cal procedure, page 75
Shows the live TEMPERATURE value on bottom line; when user presses any button the live value is
frozen and the user edits that value. If bad calibration, returns “ERR TOO LARGE TO CALIBRATE”.
Resets Conductivity calibration. After pressing
►
, select YES/NO.
Resets Temperature calibration. After pressing
►
, select YES/NO.
Enter date of calibration (MM-DD-YYYY) and initials of calibrator (XX)
CH 2 COND/RES
HOLD OUTPUTS
OFF
CH 2 COND/RES
CS
AUTO CAL ->
CH 2 COND/RES
CL CONST
MANUAL CAL ->
CH 2 COND/RES
TEMPERATURE
SET
CH 2 COND/RES
RESET CONDCAL ->
F
CH2 COND/RES
RESET TEMPCAL ->
AVERAGE
CH2 COND/RES
12-31-2015 XX
LAST CAL