GF 9950-10 Six Channel Transmitter
OPTION Menu
Common Menus
(CH 1-6) Enable AUTO SCROLL for measurement Display screens in VIEW MODE.
Set the SCROLL TIME interval for each measurement display screen.
Display available when scrolling is set to ON. Minimum 4 sec, maximum 99 seconds.
Press any key to interrupt autoscroll for 20 seconds, or the scroll time, whichever is greater.
(CH 1-6) Display Bar Graph; ON or OFF
(CH 1-6) Enter a value to represent bar at minimum.
Default = 0 (ORP = -999)
(CH 1-6) Enter a value to represent bar at maximum.
Defaults = 100 (Flow, Cond/Res, Temp), 14 (pH), 1000 (ORP), 10 (Lvl/Prs), 5 (4 to 20 mA),
42 (Sal), 20 (DO)
Function 1 - 4, Con
fi
gure the derived function, up to four functions available.
(ALL) Set the decimal to the best resolution for your application for CH 1-6.
The display will automatically scale up to this resolution for each channel.
Select 0, 1, 2, 3, or 4 decimal places, (varies by parameter).
Default = 1 Decimal Place
(pH, COND/RES, TEMP, SAL, DO only)
Set the Temperature decimal to the best resolution for your application for CH 1-6.
The display will automatically scale up to this resolution. Select 0, 1, or 2.
Default = 1 Decimal Place
(FLOW only) Set the decimal to the best resolution for the Permanent and Resettable 2 Totalizer display.
The display will automatically scale up to this resolution. Select 0, 1, 2, 3, or 4 decimal places.
Default = 1 Decimal Place
(CONDUCTIVITY/RESISTIVITY only) Set ON to automatically scale the conductivity/resistivity within range to
be displayed.
Default = OFF
(CH 1-6, FLOW only) requires password to reset the resettable totalizer.
Select OFF, ON (Does not affect Permanent Totalizer).
Default = OFF
(ALL) Select STD, CODE.
Default = STD (Refer to page 12 for further detail)
(CODE) If code is selected
Default Code = 0000 (Refer to page 12 for further detail)
CH2 BAR GRAPH MIN
GPM
0.00
CH2 BAR GRAPH MAX
GPM
100.00
DISPLAY
ON
BAR GRAPH
AUTO SCROLLING
OFF
SCROLL TIME
SEC
4
FUNCTION 1
OFF
NONE
CH1
GPM DECIMAL
_ _ _ _._
CH1
°F DECIMAL
_ _ _ _._
CH1
OFF
TOTAL LOCK
PASSWORD TYPE
%
STD
CH1
TOTAL DECIMAL
_ _ _ _._
PASSWORD TYPE
%
CODE
COND/RES
OFF
AUTORANGE