41
GF 9950-10 Six Channel Transmitter
This is the normal display and does
not time out.
DISSOLVED OXYGEN Setup Checklist (3-2610-51)
2610 wiring on page 4.
1. Make sure DISSOLVED OXYGEN sensor type is selected.
(Refer to System Setup Menu, page 11)
2. Set Units of Measurement (PPM, %SAT, TOR).
3. Set the Temperature Units (°C or °F).
4. Set Salinity reference value.
5. Set Barometric reference value.
6. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
7. Select source for Relay output (PPM or TEMP).
8. If applicable, set up relay functions for your application.
View Measurement 3 displays the primary value and the channel name. This screen does not time out.
Displays Cap Expiration Date MM-DD-YYYY. If sensor cap is missing, CH# 00/00/0000 will be
displayed.
Dissolved Oxygen
VIEW Mode Menu
CH1
5.0
DO
CH2
10.9
DO
PPM
PPM
20.00
0.00
20.00
0.00
CH1
5.0
20.0
10.9
20.0
CH2
PPM
°C
PPM
°C
20.00
0.00
20.00
0.00
CAP EXPIRATION
CH2 00/00/0000
°N
CH1 00/00/0000
F
INPUT Menu
Press ENTER to start assigning sensors to each channel.
Use the
▲
and
▼
arrow to go to any of the six channels directly.
NOTE: Display only available when the 9950 is in FACTORY mode.
Press the
►
arrow, use the
▲
and
▼
arrow to select which TYPE of sensor will be assigned to the
channel(s) (1-6).
Choose Dissolved Oxygen press ENTER to save sensor type.
Press the
►
arrow, using the
▲
or
▼
to select the input source to the channel.
Sources list will only contain Dissolved Oxygen sensors or modules.
To change sensor selection after it's been assigned, change the source to NONE, then the sensors can
be reassigned.
English, French, German, and Spanish Only. If desired, a custom name can be entered.
Enter 17-character string.
Default = DO
Set the units of measurement:
PPM = DO in mg/L; % SAT = DO % saturation; TOR = Oxygen partial pressure. Default = PPM
Manually set value to match application (0 - 42 PSU). Practical Salinity Units (1 PSU = 1g/kg, or 1 PPT)
(Parts per Thousand).
Fresh water = 0.00 PSU. Default = 0.00
SETUP
CHANNEL 2
CH 2 FACTORY
FACTORY
TYPE
CH 2 DO
DO
SOURCE
ENTER
CH2 DO
DO
AVE
NAME
AGE
CH2 DO
MEASUREMENT
PPM
CH2 DO
SALINITY (PSU)
0.0