Electromagnetic compatibility
Guidance and manufacturer’s declaration — electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. It is the responsibility
of the hospital to assure that the device is used in such an environment.
Immunity test
IEC 60601–1–2 test
level
Compliance level
Electromagnetic environment —
guidance
IEC 61000-4-2
± 15 kV air
± 15 kV air
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical Fast
Transient/ Burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 2 kV for power
supply lines
Surge IEC 61000-4-5
± 0.5 kV, ± 1 kV
differential mode
± 0.5 kV, ± 1 kV
differential mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips
IEC 61000-4-11
U
t
= 0%, 0.5 cycle (0,
45, 90, 135, 180, 225,
270, and 315 degrees)
U
t
= 0%, 1 cycle
U
t
= 70%, 25/30 cycles
(0 degrees)
U
t
= 0%, 0.5 cycle (0,
45, 90, 135, 180, 225,
270, and 315 degrees)
U
t
= 0%, 1 cycle
U
t
= 70%, 25/30 cycles
(0 degrees)
Mains power quality should be
that of a typical commercial
or hospital environment. If the
user of the equipment requires
continued operation during
power mains interruptions,
it is recommended that the
equipment be powered from an
uninterruptible power supply or a
battery.
Short interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
U
t
= 0%, 250/300
cycles
U
t
= 0%, 250/300
cycles
Mains power quality should be
that of a typical commercial
or hospital environment. If the
user of the equipment requires
continued operation during
power mains interruptions,
it is recommended that the
equipment be powered from an
uninterruptible power supply or a
battery.
Power Frequency
Magnetic Field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz and 60 Hz
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
NOTE
U
t
is the AC mains voltage prior to application of the test level.
NOTE
The EHL Platform computer must always be powered from an uninterruptible
power supply.
192
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...