Alarm messages
Alarm message
Location
Possible explanations
Suggested actions
Hub:
●
Weak Resp
connection
Central Viewer:
●
Weak Resp
connection
●
al. area
Connection between the Hub
and sensor is poor and there
was a data loss between the
Hub and the sensor.
●
Place the Hub closer to the
patient.
●
If the problem persists,
contact qualified service
personnel.
Hub:
●
Resp connection lost
●
Connection lost
Central Viewer:
●
Resp connection lost
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
Connection between the Hub
and sensor was lost due to too
long a distance between them
(for instance, the patient has
walked away from the Hub).
●
Check the distance between
the sensor and the Hub.
●
Make sure the patient
understands how far they
can move from the Hub
without losing the wireless
connection.
●
If the problem persists,
contact qualified service
personnel.
Hub:
●
Resp device
overheating
●
Device overheating
Central Viewer:
●
Resp device
overheating
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
Sensor battery surface
temperature is too high.
The maximum surface
temperature of the devices
being worn by the patient (Hub,
sensor battery and wearable
sensors) will never exceed
43°C (The device will shut
down before the temperature
reaches 43°C). When the surface
temperature is approaching
43°C, an overheating alarm
is given, allowing the clinician
to take appropriate action to
ensure the time of exposure is
limited based on the condition
of the patient.
●
Check the sensor battery and
ensure that it is not covered
by anything that might lead
to overheating.
●
If the problem persists,
replace the sensor battery.
Alarms related to the SpO
2
and pulse rate
measurement
Alarm priorities are given in the section discussing alarm specifications.
Make sure that you are familiar with the generic layout of the Hub and Central Viewer
screens. This will help you identify where the following alarm messages appear. The alarm
location is indicated with the following abbreviations:
●
al. area = alarm area
●
data = data area
5815148-01
Portrait
™
Mobile Monitoring Solution
139
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...