Impedance respiration
In addition to true breaths, patient motion can also change the impedance of the
chest, and is thus visible in the impedance waveform. Motion artifact can be several
times larger in amplitude than normal breathing.
The following graphic illustrates how motion artifact is reflected in the respiration
waveform:
In this example:
●
1–11 = room-bed identification
●
Smith, Peter = patient’s last name, first name
●
RR/min (
RR/min
) = respiration rate label and unit of
measure
The top row below the RR value 20 shows the waveform
during breathing. At the end of the top row, and during
the entire second row, the waveform shows large artifacts
caused by a moving patient. Patient motion reduces the
RR value reliability, which is indicated by the SQI indicator
having only one filled circle.
Waveform scale is indicated on the top row (in this
example: 2x).
Respiration measurement practicalities
Instruct the patient regarding the following issues. Always consider your hospital
guidelines as well.
●
The respiration sensor should be detached before having a shower. The respiration
patch can be left attached.
■
After the shower:
♦
Check that the electrodes are properly attached to the skin and that there is
no skin irritation. If the electrodes are no longer properly attached or the
adhesive no longer holds, replace the respiration patch.
♦
Make sure that the respiration patch connector is completely dry before
reconnecting the sensor.
Respiration points to note
●
This equipment is suitable for use in the presence of electrosurgery, as tested
according to IEC 80601-2-49 clause 202.8.102 Electrosurgery interference.
●
Defibrillator discharge may affect the respiration measurement. Defibrillation
recovery time after discharge is ≤15 seconds.
●
Materials used in accessories are not toxic. Accessories do not contain natural
rubber latex.
●
Preferably use the measurement sites as indicated in the instructions delivered
with the respiration patch to ensure reliable and accurate measurement.
5815148-01
Portrait
™
Mobile Monitoring Solution
95
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...