Pulse oximetry and pulse rate
1.
Press both sides of the probe gently to make the probe more rounded. This will
help when inserting the finger.
2.
Align the probe with the GE logo upwards and insert the finger all the way to
the end of the probe.
Using the P-W01 and P-SE01 sensors
These P-W01 and P-SE01 sensor-specific instructions explain the use in more detail.
Also follow the generic instructions for starting the SpO
2
measurement.
Note the following:
●
The sensor tape and foam are for single-patient use only.
●
Sensitivity to the adhesive tape may cause an allergic reaction.
1.
Slide the probe through the slits in the wrap. Alternatively, align the star on the
tape with the light source.
2.
Ensure a good contact between the probe and the measurement site. Do not
restrict circulation.
To replace the wrap:
1.
Gently remove the worn wrap.
2.
Clean the light source and detector.
3.
Place the sensor in a new wrap.
Checking the SpO
2
and pulse rate measurement
Perform the following checks to ensure that the measurement is working.
1.
On the Hub:
a. Check that the wireless network connection LED on the Hub is lit and keeps
alternating between dark and light green.
b. Check that the pulse search animation (three dots in a horizontal line) and the
message
Pulse search
are displayed. When they disappear, you can see the
measurement parameter (SpO
2
, PR) on the Hub screen.
c. Check the measurement values and the signal quality indicator to ensure
measurement quality.
2.
On the Central Viewer:
a. Check that the parameter waveform and signal quality indicators are
displayed in the Single Patient View and the parameter values in the Single
and Multiple Patient Views.
5815148-01
Portrait
™
Mobile Monitoring Solution
105
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...