Technical specifications
Design, environmental, and physical
specifications
For additional specifications, refer to the documentation provided with your
equipment. For more information on products that can be used with the system,
refer to the list of Compatible devices.
Operating altitude: The system shall meet specifications when subjected to altitudes
corresponding to pressure readings from 620 mbar to 1060 mbar.
NOTE
An average pressure reading of 620 mbar corresponds to an
altitude of 3950 m (12,959 ft) and 1060 mbar corresponds to
an altitude of –380 m (–1247 ft).
Non-operating altitude: The system shall meet specifications after being subjected to
altitudes corresponding to pressure readings from 500 mbar to 1060 mbar.
NOTE
An average pressure reading of 500 mbar corresponds to an
altitude of 5486 m (18,000 ft).
WARNING
INACCURATE READINGS. After transport or storage of the
device outside the specified operating temperature range,
always allow the device to stabilize back to operating
temperature range before applying power to it. Using the
device outside the specified operating environment may
result in inaccurate results.
FCC information
●
Hub FCC ID: 2AO8L-HUB01; contains FCC ID: 2AO8L-WL18DBMOD
●
Sensor battery FCC ID: 2AO8L-SBT01
This device complies with part 15 and Part 95 of the FCC Rules. Operation is subject
to the following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including interference that
may cause undesired operation.
CAUTION
UNEXPECTED SYSTEM PERFORMANCE. Changes or
modifications not expressly approved by the party responsible
for compliance could void the user’s authority to operate
the equipment. This transmitter must not be co-located or
operated in conjunction with any other antenna or transmitter.
5815148-01
Portrait
™
Mobile Monitoring Solution
145
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...