C
Default settings and configuration
About configuration settings
There are certain system default configuration settings that are available for an
authorized clinical user. These settings can be found in Config UI >
Configuration
>
Default Profiles
>
Patient
, and they are password-protected.
Certain settings have different defaults per
Care profile
(
Age 3-4
,
Age 5-6
,
Age 7-12
,
Age 13-17
,
Adult
). Changes to the patient default profiles are not available to already
admitted patients. The changes made will be applied to the next admitted patient
after the changes have been saved.
The Config UI includes many other settings as well, but they are only available with
service access rights.
WARNING
PATIENT SAFETY. In case the Hub does not receive a
configuration from the Core Services for some reason, it will
use the factory defaults. Always check the settings to ensure
patient safety.
Config UI Patient default profiles configuration
Navigate to Config UI >
Configuration
>
Default Profiles
>
Patient
.
NOTE
When the system is first installed, the care profile defaults
in the patient default settings are customized for the
age-specific profiles. Selecting
Reset All
for a care profile will
reset them to the common
Facility
-level settings. If you wish
to revert them back to the factory default settings, you need
to enter those factory default values manually.
Setting
Description
Unit of
measure
Options
Default value
Alarm settings
Enable timed alarm
acknowledgement
n/a
Yes
or
No
Yes
Alarm acknowledgement time
Not adjustable.
Duration for
timed alarm
acknowledge-
ment.
Seconds
n/a
120
5815148-01
Portrait
™
Mobile Monitoring Solution
177
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...