Managing patients and monitoring
1.
Insert the Hub into the pouch. Check that the Hub parts indicated with the
exclamation mark are aligned with the pouch as in the graphic.
It is especially important that the speaker is not covered by the pouch to avoid
missing any audible alarms.
2.
Instruct the patient that they can wear the Hub pouch when moving around with
the Hub. The Hub pouch can be worn in different ways: neck, shoulder, back etc.
If the pouch or Hub get wet, dry them to ensure correct functioning of the speaker
and touchscreen. Always replace a dirty or broken Hub pouch.
Unassigning the patient from the Hub
Unassigning a patient from a Hub ends the monitoring, deletes the patient data from
this Hub, and clears the sensor(s) and Hub pairing. It also allows the Hub to be used
for another patient. You can do this at either the Hub or the Central Viewer.
1.
Unassign the patient using the Hub:
a. Activate the screen and enter the passcode.
b. Tap the patient name to access the patient detailed view.
c. Select
End monitoring
.
A message appears asking you to confirm that you wish to end monitoring.
d. Press
to unassign the patient from the Hub, or select the X on the
upper right corner of the message to return to the previous view without
clearing the Hub and patient assignment.
The assignment between the patient and the Hub is cleared. The message
Patient
data removed
appears. The patient stays admitted to the monitoring system until
discharged from the Central Viewer.
2.
Unassign the patient from the Hub using the Central Viewer:
a. Select a patient slot in the Multiple Patient View.
b. From the Single Patient View action menu, select
Remove Hub
.
The
Remove Hub?
menu appears.
c. Select
Yes, Remove
to unassign the patient from the Hub, or select
Cancel
to
return to the previous screen without clearing the Hub and patient assignment.
After selecting
Yes, Remove
, the assignment between the patient and the Hub
is cleared and the devices can be used for a different patient after cleaning and
disinfection. The patient stays admitted to the monitoring system until discharged
from the Central Viewer.
Shutting down the Hub
Make sure that the Hub is not needed for monitoring a patient before shutting it down.
The Hub is shut down with the power key (1.) located on the side of the Hub:
70
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...