Default settings and configuration
Setting
Description
Unit of
measure
Options
Default value
SpO2 low - alarm priority
Select the
priority for the
SpO2 low
alarm.
If you set the
priority value
so low that it
deviates from
the recommen-
dation of inter-
national alarm
safety stan-
dards, the value
will be accom-
panied by
and a warning
message ap-
pears.
n/a
High| Medium|
Low| Info
Medium
SpO2 low - alarm delay
Select the delay
after which the
SpO2 low
alarm
is triggered.
If the
SpO2
critically
low
alarm is
triggered before
the
SpO2 low
alarm delay has
expired, these
two alarms
are triggered
simultaenously.
Seconds
5, 10, 15, 20, 30,
60, or 90
Age 3-4
: 20
Age 5-6
: 20
Age 7-12
: 30
Age 13-17
: 60
Adult
: 60
Facility
: 20
SpO2 critical low (% points below
set SpO2 low limit)
Delta setting
that should be
used against
the
SpO2
low
limit for
determining the
critical low limit
setting.
Percent
1 to 15
Age 3-4
: 3
Age 5-6
: 3
Age 7-12
: 3
Age 13-17
: 6
Adult
: 6
Facility
: 3
SpO2 critical low - alarm priority
Not adjustable.
This is the
priority for the
SpO2 critically
low
alarm.
n/a
n/a
High
PR high - alarm on
Not adjustable.
n/a
n/a
Yes
180
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...