Graphical user interface
●
Alarm limits
●
Waveforms
●
Signal quality indicators for each displayed real-time numeric value
●
Trends with event markers
●
A list of active events
Only the parameters currently being monitored for a patient appear in the Single
Patient View.
●
Main header area: Displays at all times. Displays the care area name, the time
in 24-hour format (HH:MM), system messages related to the Central Viewer, and
volume control.
●
Alarm display unit area: Displays alarms for other patients in the care area.
●
Patient header: Displays the room number, bed number (if applicable), patient
name, up to two patient identifiers, sex, date of birth, pacemaker information (
Yes
with
indicator,
No
,
Unknown
with
indicator) and Hub identifier. Provides
access to actions menu.
●
Active alarms area: Displays the active alarms for this patient (highest priority, most
recent unacknowledged alarm at the top of the list).
●
Live data area: Displays the real-time data for the patient, including waveforms,
limit configurations, signal quality indicator, and parameter numerics.
●
Trends area: Displays the trended data for the individual patient for the previous
2, 4, 8, 12, or 24 hours
To return to the Multiple Patient View, close the Single Patient View by selecting the X
indicator in the upper right corner of the screen.
Alarm display unit area
An alarm display unit area is shown at the top of the Single Patient View to provide
visibility to active alarms on other patients configured to the same Central Viewer.
Selecting an alarm message opens the Single Patient View for the alarming patient.
An alarm message is displayed in the alarm display unit area when there is an active,
unacknowledged alarm of high or medium priority for another patient. Only one alarm
message is displayed per patient; if there are multiple alarms for a patient, only the
most important (highest priority, most recent unacknowledged) alarm is displayed. If
there are more than one alarms of the same importance, only the most recent one of
those is displayed. An alarm message remains visible until the alarm is acknowledged
or the triggering condition clears.
Alarm messages are sorted from left to right by priority and then by time. For instance,
two high priority alarms and two medium priority alarms would be shown as:
●
First alarm from the left = the most recent high priority alarm
●
Second alarm from the left = a previous high priority alarm
●
Third alarm from the left = the most recent medium priority alarm
●
Fourth alarm from the left = a previous medium priority alarm
A maximum of six alarm messages are displayed at once.
The alarm display unit area contains the following information:
●
The priority of the alarm is indicated by the color of the alarm display unit area.
High priority alarms are red, medium priority alarms are yellow.
54
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...