Default settings and configuration
Setting
Description
Unit of
measure
Options
Default value
Alarm volume - Hub
Percent
0, 10, 20, 30, 40,
50, 60, 70, 80,
90, 100
100
Mode settings
Hub monitoring mode
n/a
Local| Remote
Local
Resp limit alarms
RR high - alarm on
Not adjustable.
n/a
n/a
Yes
RR high - alarm limit
Select the
threshold alarm
limit for the
RR
high
alarm.
Breath/Min
4 to 80
Age 3-4
: 40
Age 5-6
: 30
Age 7-12
: 30
Age 13-17
: 30
Adult
: 30
Facility
: 30
RR high - alarm priority
Select the
priority level
for the
RR high
alarm.
n/a
High| Medium|
Low| Info
Low
RR high - alarm delay
Select the delay
after which the
RR high
alarm
is triggered.
Seconds
15, 30, 90, or 120
Age 3-4
: 15
Age 5-6
: 15
Age 7-12
: 15
Age 13-17
: 15
Adult
: 30
Facility
: 15
RR low - alarm on
Not adjustable.
n/a
n/a
Yes
RR low - alarm limit
Select the alarm
limit for the
RR
low
alarm.
Breath/Min
4 to 80
Age 3-4
: 15
Age 5-6
: 15
Age 7-12
: 10
Age 13-17
: 8
Adult
: 8
Facility
: 15
RR low - alarm priority
Select the
priority level
for the
RR low
alarm.
n/a
High| Medium|
Low| Info
Medium
RR low - alarm delay
Select the delay
after which the
RR low
alarm is
triggered.
Seconds
5, 10, 15, 20, or
30
Age 3-4
: 5
Age 5-6
: 5
Age 7-12
: 10
Age 13-17
: 10
Adult
: 15
Facility
: 5
Apnea - alarm on
Not adjustable.
n/a
n/a
Yes
178
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...