Cleaning, disinfection, and care
2.
Wring excess liquid from the cloth (wet but not dripping).
3.
Wipe the exterior surface. Ensure that the entire surface is wet and make sure all
surfaces are uniformly cleaned.
●
Pay special attention to hard-to-clean areas (marked here with red), like
grooves and crevices):
●
Use cotton swabs to access hard to reach areas. Do not use any sharp tools to
clean the device.
4.
Replace soiled and/or dried wipes or non-linting cloths as needed.
5.
Repeat the cleaning until the device is visibly clean.
6.
Discard disposable wipes or non-linting cloths according to your hospital
guidelines. Do not reuse wipes.
7.
Rinse the device thoroughly using a soft, clean non-linting cloth saturated with
lukewarm tap water (temperature range 27 to 44°C, 81 to 111°F).
8.
Allow the device to air dry until it is visually dry, or dry with thoroughly with a soft,
clean, non-linting cloth before placing the device into the charger.
Drying times may vary based on the environmental conditions. Do not let fluid
pool around connection pins to avoid corrosion. If this happens, blot dry with a
cotton swab or soft cloth.
●
Pay special attention to metal parts (marked here with red) when drying device
to avoid corrosion:
Post-cleaning inspection
Inspect devices visually after cleaning and prior to disinfection for the following:
●
Cleanliness
■
If there is any visible soil present, repeat the cleaning procedure until no visible
soil remains.
●
Damage, including, but not limited to, corrosion (rust, pitting), discoloration,
excessive scratches, flaking, cracks, and wear
●
Missing or worn device labeling or marking
Do not use devices that are not functioning properly or are damaged, or have
unrecognizable markings, or missing or worn device labeling or marking.
5815148-01
Portrait
™
Mobile Monitoring Solution
123
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...