Alarm specifications
lost,
Network connection lost
alarms are generated on the Hub and on the Viewer
(delays reported below).
The determination of an alarm condition has nominal factors (such as signal quality
and movement artifact) that may affect the ability to derive the parameter value.
The following disclosures list the alarm system delays as an average and include the
delay of the first activated priority alarm.
Alarms with alarm delays of more than 10 s are specified with the average, minimum,
and maximum values obtained from five trial measurements.
NOTE
Any alarm delays that are less than 5 s are given as <5 s.
Alarm delay specifications for impedance respiration
alarms
The following table lists the alarm delays for physiological alarms related to the impedance
respiration measurement. Any alarm delays that are less than 5 s are given as <5 s.
Physiological alarms
Alarm
Limit range
Limit increment
Alarm delay
Apnea
10 to 60 s
1 s
32 s (average) (32 to 32 s range)
RR high
4 to 80 breaths
per minute
1 breath per
minute
32 s (average) (32 to 33 s range)
RR low
4 to 80 breaths
per minute
1 breath per
minute
24 s (average) (24 to 25 s range)
Technical alarms
Alarm
Alarm delay
Change Resp battery
50 s (average) (45 to 57 s range)
Check Resp patch
30 s (average) (30 to 31 s range)
Clean Resp battery interface
<5 s
No Resp battery
<5 s
Resp battery low
48 s (average) (41 to 55 s range)
Resp connection lost
10 s (average) (9 to 10 s range)
Resp device overheating
<5 s
Weak Resp connection
<5 s
5815148-01
Portrait
™
Mobile Monitoring Solution
153
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...