20
Alarm specifications
Auditory alarm volume, IEC tones
Tested in accordance with IEC 60601-1-8 subclause 6.3.3.2 with alarm volume control set to
maximum level (100%) and minimum level (10%)..
Alarming unit
Alarm volume setting
Maximum sound pressure level
100%
High priority alarm: 68 dB(A)
Medium priority alarm: 67 dB(A)
Low priority alarm: 56 dB(A)
70%
High priority alarm: 61 dB(A)
Medium priority alarm: 60 dB(A)
Low priority alarm: 52 dB(A)
10%
High priority alarm: 46 dB(A)
Medium priority alarm: 42 dB(A)
Low priority alarm: 37 dB(A)
Hub
Alarm volume setting
Maximum sound pressure level
100%
High priority alarm: 63 dB(A)
Medium priority alarm: 61 dB(A)
Low priority alarm: 50 dB(A)
70%
High priority alarm: 54 dB(A)
Medium priority alarm: 53 dB(A)
Low priority alarm: 39 dB(A)
10%
High priority alarm: 36 dB(A)
Medium priority alarm: 36 dB(A)
Low priority alarm: 27 dB(A)
Audio alarm sound tolerances
Tolerances for sounds (total pulse duration with rise and fall times) are:
●
Time < 5 s: ±5%
5815148-01
Portrait
™
Mobile Monitoring Solution
151
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...