Alarms
Alarm signals and user position
Alarm signals are designed to alert the user visually and audibly. The signals should be
detectable up to a distance of 1 meter/3.3 feet from the alarming device.
Visual alarm indicators
Visual alarm indicators have the following characteristics:
Priority level
Alarm location
High
Medium
Low
Informational
Characteristics
Hub alarm area,
Central Viewer
alarm message
area (Single
Patient View)
White text
inside a red box.
Black text inside
a yellow box.
Black text inside
a cyan box.
Black text inside
a gray box.
Only alarms for
the displayed
patient appear.
The alarms are
displayed until
the condition is
resolved.
Hub alarm area,
Central Viewer
alarm message
area (Multiple
Patient View)
White text
inside a red box.
Black text inside
a yellow box.
Black text inside
a cyan box.
Black text inside
a gray box.
The highest
priority, most
recent unac-
knowledged
alarm is dis-
played until the
condition is re-
solved.
Central Viewer
only: alarm
display unit area
(Single Patient
View)
White text
inside a red box.
Black text inside
a yellow box.
No indication
(normal color).
No indication
(normal color).
Other patients
than the one
displayed in this
slot: Only the
highest priority,
most recent
alarm from
each alarming
patient is
displayed in the
alarm message
area until the
condition is
resolved or
the alarm is
acknowledged.
Hub and
Central Viewer:
Measurement
numerics
White text
inside a red box.
Black text inside
a yellow box.
Black text inside
a cyan box.
No indication
(normal color).
The
measurement
numeric
displays in
the alarming
state until the
condition is
resolved. A
measurement
numeric value
that is related
to an alarm
is highlighted
accordingly.
84
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...