Safety and intended use
WARNING
UNEXPECTED SYSTEM PERFORMANCE. Compatibility is critical
to safe and effective use of this device. Only use components
that are listed as part of the system and as compatible. Using
incompatible components can result in loss of monitoring. For
more information, see the list of compatible devices.
WARNING
UNEXPECTED SYSTEM PERFORMANCE. External radiating
devices (e.g. cell phones, brachytherapy devices, x-ray
devices) may disturb the measurement. Do not use such
devices close to the system.
WARNING
UNEXPECTED SYSTEM PERFORMANCE. Radio-frequency
identification (RFID) readers that operate at frequencies
134.2 kHz or13.56 MHz may negatively impact the system
performance. Do not use such devices close to the system.
EMC warnings
WARNING
DEGRADED PERFORMANCE. Do not use portable RF
communications equipment (including peripherals such as
antenna cables and external antennas) closer than 30 cm (12
inches) to any part of this device, including cables specified
by the manufacturer. Otherwise the performance of this
device/system may degrade.
WARNING
DEGRADED PERFORMANCE. Do not use an RFID reader
operating at a low frequency such as 134.2 kHz and 13.56
MHz close to any part of the Hub and sensors. Otherwise
there is a risk of degraded performance of the impedance
respiration measurement.
●
Impedance respiration measurement value may be out
of the measurement accuracy range and respiration
waveform noise may be randomly high when the
respiration sensor is too close to such RFID readers. When
these RFID emissions are removed, the measurement
returns to expected measurement accuracy within 60
seconds without any changes to settings.
WARNING
ELECTROMAGNETIC INTERFERENCE. Magnetic and electrical
fields are capable of interfering with the proper performance
of the device. For this reason make sure that all external
devices operated in the vicinity of the device comply with
the relevant EMC requirements. X-ray equipment or MRI
devices are a possible source of interference as they may
emit higher levels of electromagnetic radiation. Changes or
modifications to this device/system not expressly approved
by GE may cause EMC issues with this or other equipment.
This device/system is designed and tested to comply with
applicable standards and regulations regarding EMC and
needs to be installed and put into service according to the
EMC information stated as follows: This device/system is
suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes. Mains power should be that of a typical commercial
or hospital environment. Device is compliant to Class A.
24
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...