Cleaning, disinfection, and care
●
Visually ensure that no substance residue remains on the device.
Permitted disinfectants
All third party trademarks are property of their respective owners.
Trademark names and product availability may vary by country.
●
CaviWipes wipes
●
Super Sani-Cloth (purple top)
Disinfection procedure
Ensure that a thorough cleaning of the device has been performed prior to disinfection
Charger: Unplug the power cord from the wall outlet.
Disinfecting the devices
The following disinfection instructions apply to the Hub, SpO
2
and respiration sensors,
sensor batteries, and charger. Any device-specific instructions are indicated.
1.
Apply disinfectant solution to a clean non-linting cloth or use approved
disinfectant wipes.
2.
Wipe the entire device surface. Treated surfaces must remain visibly wet for a
time indicated in the disinfectant manufacturer’s documentation.
3.
Apply additional disinfectant solution or use additional fresh disinfectant
wipes to ensure that the device remains wet continuously per the disinfectant
manufacturer’s instructions for contact time.
●
Pay special attention to hard to reach areas like recessed areas and ridges. To
identify these, refer to the graphics in the cleaning instructions. To disinfect
them, dampen a cotton swab with the disinfectant solution.
●
Hub: Do not use any sharp tools when disinfecting the device.
4.
Discard disposable wipes or non-linting cloths according to your hospital
guidelines. Do not reuse wipes.
5.
Manually rinse the device by wiping it for 30 seconds using a sterile, soft,
non-linting cloth saturated with purified water (PURW) (temperature range 27
to 44°C, 81 to 111°F).
a. Repeat two additional times for a total of three rinse cycles, using fresh
purified water and a new sterile cloth for each cycle.
6.
Dry the device:
a. Hub, sensors, sensor batteries: Dry using a sterile, non-linting cloth until the
device is visually dry before placing it into the charger or taking it into use
b. Charger: Dry using a sterile, non-linting cloth until the device is visually dry
before taking it into use
7.
Charger: connect the charger back into a wall socket when it is dry.
5815148-01
Portrait
™
Mobile Monitoring Solution
127
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...