System description
Introduction to the alarming unit
The alarming unit is connected to the Central Viewer computer with a USB cable. This
cable is integrated in the alarming unit and cannot be removed.
WARNING
PERSONAL INJURY. Route the cable of the alarming unit in
such a way that it will not present a stumbling hazard.
There is one LED indicator on the front of the alarming unit:
The LED indicates the following:
●
Flashing green: the device is powered up and waiting to authenticate the
application (for example when the computer is on, but the application is not up
and running).
●
Solid green: the device has been authenticated and is communicating with the
Central Viewer.
●
Flashing yellow: the device has not been authenticated and is not communicating
with the Central Viewer. This can happen, for instance, when the Central Viewer
application is turned off erroneously or the computer has lost power. The yellow
LED also flashes if there is a fault in the alarming unit itself. In all of these cases the
audible power failure alert is active. If necessary, you can silence the audible power
failure alert by pressing the button (1) on the back of the alarming unit. This will
disable the power failure alarm until the computer power is restored.
1. Audio on/off button
For all service and maintenance instructions and hardware-related issues, refer to
the service manual.
44
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...