Electromagnetic compatibility
SpO
2
essential performance when signal strength is indicated as normal ( ) and no messages about
questionable signal quality are displayed
Pulse oximetry saturation measurement value
accuracy (a root-mean-square difference for
adult and pediatric patients per ISO 80601-2-61,
Clause 201.12.1.101.1)
SpO
2
(70 to 100%): ±2
Pulse oximetry peripheral pulse rate accuracy
(a root-mean-square difference for adult and
pediatric patients)
PR (30 to 250 bpm): ±2 bpm
Electromagnetic emissions
Guidance and manufacturer’s declaration — electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the
user of the device should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Portrait
TM
CVA01, CAUO1, and HUB01 use RF energy only for
their internal function.
Therefore, RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
The equipment is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
NOTE
The emissions characteristics of this equipment make it
suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might not
offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
Electromagnetic immunity
Guidance and manufacturer’s declaration — electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. It is the responsibility
of the hospital to assure that the device is used in such an environment.
Immunity test
IEC 60601–1–2 test
level
Compliance level
Electromagnetic environment —
guidance
Electrostatic Discharge
(ESD)
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
Floors should be wood, concrete
or ceramic tile. If floors are
5815148-01
Portrait
™
Mobile Monitoring Solution
191
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...