Alarm messages
Technical messages
Alarm message
Location
Possible explanations
Suggested actions
Hub:
●
Pulse search
●
al. area
SpO
2
and PR values are not yet
available due to ongoing signal
control and data acquisition.
●
Reposition the probe.
●
Clean the application site
(remove nail polish, for
instance).
●
If the problem persists,
contact qualified service
personnel.
Hub:
●
Check SpO2 probe
●
Check probe
Central Viewer:
●
Check SpO2 probe
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
Prolonged pulse search.
The signal is too noisy for
measuring SpO
2
.
Peripheral perfusion is too low
for measuring SpO
2
.
●
Check the sensor and
connections.
●
Check the patient.
●
Isolate noise sources.
●
Change the measurement
site.
Hub:
●
SpO2 probe off
●
Probe off
Central Viewer:
●
SpO2 probe off
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
SpO
2
probe is off the patient.
●
Check the connections and
adjust as needed.
Hub:
●
Faulty SpO2 probe
●
Faulty probe
Central Viewer:
●
Faulty SpO2 probe
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
SpO
2
probe is faulty (open or
short circuit failure).
●
Check the probe and replace
if necessary.
Hub:
●
SpO2 battery low
Central Viewer:
●
SpO2 battery low
●
al. area
Sensor battery charge is low.
●
Charge the battery.
●
Replace the battery.
Hub:
●
Change SpO2 battery
Central Viewer:
●
Change SpO2 battery
●
al. area
Sensor battery is empty.
●
Replace the battery and
put the empty one into the
charger.
Hub:
●
Clean SpO2 battery
interface
●
Clean battery
interface
Central Viewer:
●
Clean SpO2 sensor
interface
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
There is dirt in the sensor.
●
Check and clean the sensor.
5815148-01
Portrait
™
Mobile Monitoring Solution
141
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...