Cleaning, disinfection, and care
2.
If the Hub is not in clinical use, make sure that it is powered off. If the Hub is
powered on:
a. Activate the screen by pressing the
key and entering the passcode
with the on-screen keyboard. Confirm with the
key.
b. Press the power key for 1.5 seconds.
When the Hub screen is facing you, the power key is located on the right hand
side of the Hub.
c. Confirm that you wish to continue with the shut down.
3.
If the Hub is in clinical use, make sure that the screen is locked (you can see
texts
No patient
or
No monitoring
, or only the current date and time (
REMOTE
monitoring mode), or measurement values with the lock indicator
at the bottom
of the screen (
LOCAL
monitoring mode). If the Hub is not locked, lock it by pressing
the power key.
4.
Visually ensure that the device is intact. Evidence of damage may include, but is
not limited to, discoloration, excessive scratches, wear, and cracks.
Cleaning the Hub
In case there is some dried or gross soil, perform pre-cleaning by wiping with a soft,
clean non-linting cloth saturated with lukewarm tap water (temperature range 27 to
44°C, 81 to 111°F) until visible dried or gross soil is removed.
1.
Dampen a soft non-linting cloth with detergent or use ready-to-use detergent
wipes.
Prepare the detergent according to the manufacturer’s instructions.
2.
Wring excess liquid from the cloth (wet but not dripping).
3.
Wipe the exterior surface. Ensure that the entire surface is wet and make sure all
surfaces are uniformly cleaned.
●
Pay special attention to hard-to-clean areas (marked here with red), like
grooves and crevices:
Use cotton swabs to access hard to reach areas. Do not use any sharp tools to
clean the device.
4.
Replace soiled and/or dry wipes or non-linting cloths as needed.
5.
Repeat the cleaning until the device is visibly clean.
6.
Discard disposable wipes or non-linting cloths according to your hospital
guidelines. Do not reuse wipes.
116
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...